Customization: | Available |
---|---|
Classification: | Defibrillator |
Type: | First-Aid Devices |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Defibrillator Output
|
Biphasic Truncated exponential
|
Energy Sequence(child)
|
50J,50J,50J
|
Energy Sequence(Adult)
|
150J,200J,200J
|
Charge Time
|
6 sec. to 150J
8 sec. to 200 J(New battery) |
Analysis Time
|
6 sec.
|
The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery
|
Less than 10 seconds
|
The maximum time from the initiation of rhythm analysis to readiness for discharge after 6 shocks
|
Less than 10 seconds
|
The maximum time from initially switching power on toreadiness for discharge
|
Less than 18 seconds
|
Visual Prompts
|
LED prompts;Automatic Self-test
|
Controls Output Energy Accuracy
|
Four buttons On/Off, ShockAdult/childCPR
±10% into any impedance from 25 to 175Q |
The Maximum Voltage
|
1400±50V
|
Output disabled when the patient impedance is outside limits
|
20Ω to 200Ω
|
Battery Part No .
|
CR123A-4x2
|
Non-rechargeable
|
2.8 Ah
|
Capacity
|
120 discharges at 200 Joules
|
Shelf Life (25°C+15C)
|
3 years standby (after installation)
|
Physical Dimensions , Weight
|
280 x 209 x 80 mm , 2.5 kg
|
Operating Temperature
|
0°C to 40°C
|
Operating Humidity
|
Relative humidity between 30% and 95%(non-condensing)
|
Storage Temperature (without battery)
|
-20°C to 55°C
|
Storage Humidity (without battery)
|
Up to 93%(non-condensing)
|