Customization: | Available |
---|---|
Customized: | Customized |
Certification: | CE |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Quick Details: |
10.1 inch high resolution color TFT display |
Three Working Modes: Monitoring, Surgery and Diagnosis. |
Standard 6 parameters configuration: ECG, RESP, TEMP NIBP SPO2, PR |
Bult-in high capacity battery up to 3 hours of continuous work |
Multi-display modes suitable for different applications Standard Interface, Large font, ECG Standard Full Display, OXY,Trend Table, BP Trend. |
Standard Configuration | Option Configuration |
Main machine 1 PC | ETCO2, Dual IBP, Nellcor/Masimo SPO2 |
ECG lead wire 1 PC | Suntech NIBP, Anesthetic gas module, Printer |
SPO2 sensor 1 PC | Touch screen function |
1BP cuff 1 PC | Center monitoring system |
SPO2 extension cord 1 PC | Trolley, Wall mount |
BP extension cord 1 PC | 12 lead ECG |
Power cord 1 PC | |
Temperature probe 1 PC | |
User Manual 1 PC |
ECG
|
SPO2
|
NIBP
|
v Lead Mode: 5 Leads (I, II, III, AVR, AVL,AVF, V)
|
Measuring Range: 0 ~ 100 %
|
Method: Oscillometry
|
Gain: 2.5mm/mV, ´5.0mm/mV, 10mm/mV, ´20mm/mV
|
Resolution: 1%
|
Measure mode: Manual, Auto, STAT
|
Heart Rate: 15-300 BPM (Adult); 15-350 BPM (Neonatal)
Resolution: 1 BPM Accuracy: ±1% Sensitivity >200 uV(Peak to peak) |
Accuracy: 70% ~ 100% (±2 %)
|
Unit: mmHg, kPa
|
ST measurement range: -2.0 ~ +2.0 mV
Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, which is greater Other Range: unspecified Sweep speed: 12.5 mm/s, 25mm/s, 50mm/s |
Pulse Rate: 20-250 BPM
|
Measure and alarm range:
Adult Mode: *SYS 40 ~ 280 mmHg * DIA 10 ~ 225 mmHg * MEAN 20 ~ 240 mmHg Pediatric Mode: * SYS 40 ~ 220 mmHg * DIA 10 ~ 160 mmHg * MEAN 20 ~ 170 mmHg Neonatal Mode: * SYS 40 ~ 135 mmHg * DIA 10 ~ 100 mmHg * MEAN 20 ~ 110 mmHg |
Bandwidth:
* Diagnostic: 0.05~130 Hz * Monitor: 0.5~40 Hz * Surgery: 1~20 Hz |
Resolution: 1 BPM
|
Resolution: 1mmHg
|
|
Accuracy: ±3 BPM
|
Accuracy: ±5mmHg
|
RESP
|
TEMP
|
Method: Impedance between RA-LL or RALA
|
Measure and Alarm Range: 0 ~ 50 °C
|
Measurement Range:
*Adult: 10 ~ 120 BrPM
*Neonatal / Pediatric: 7 ~ 150 BrPM
|
Resolution: 0.1°C
|
Resolution: 1 BrPM
|
Accuracy: ±0.1°C
|
Accuracy: ±2 BrPM
|
|